HP 3457A Service Manual page 83

Hide thumbs Also See for 3457A:
Table of Contents

Advertisement

= _t
-_
=
Operational Verification Tests
3457A
2-Wire Ohms Operational Verification Tests Cont'd
SENSE
INPUT
TAST a
im
fri
FILTER
CIC
Figure 4-16. 2-Wire Ohms Operational Verification Test Connections
4. If any of the offset readings are beyond the limits specified in Table 4-25 and on the Test Record,
the instrument should be calibrated. Refer to Section V of this manual for calibration procedures.
5. Remove the short from the front panel input terminals.
34570
34570
34578
90 day
limits
1 year
limits
Input
Range
| Set Up
Short
3 Kohm | OHM
0.000207 Kohm | 0.000207 kohm
Short
30 Kohm | OHM
00.00027 Kohm | 00.00027 Kohm
Short
|300 Kohm | OHM
000.0010 Kohm | 000.0010 Kohm
Table 4-25. 2-Wire Ohms Offset Test Limits
4-82. 2-Wire Ohms Function - Gain Test
4-83.
Equipment
Required.
Resistance
Standards
of
3 Kohms
(+ .001%),
300 Kohms (+ 001%) are required for this procedure.
30 Kohms
(+ 001%),
and
LSet the HP 3457A
to the 2-wire ohms function (OHM) and the number of digits displayed (DIGITS
DISP) to six.
2. Connect the Resistance Standard to the HP 3457A front panel HI and LO input terminals. (The
connecting wires should be as short as possible to reduce lead resistance).
3. Use the Test Record to record the 3 Kohm, 30 Kohm and 300 Kohm full-scale readings.
4. If any of the readings are beyond the limits specified in Table 4-26 and on the Test Record, the in-
strument should be calibrated. Refer to Section V of this manual for calibration procedures.
5. Remove the Resistance standard from the HP 3457A front panel input terminals.