HP 3457A Service Manual page 111

Hide thumbs Also See for 3457A:
Table of Contents

Advertisement

PERFORMANCE TEST CARD
1 YEAR LIMITS
Hewlett-Packard Model 3457A
Digital Mutimeter
Serial Number
2-WIRE OHMS TEST
Test Performed by
Date
Reference Temperature
Step#
Input
to
Set-Up
and
High
Reading
Low
Test
Test
34578
Configuration
Limit
Limit
Pass
Fail
OFFSET
TEST
1
Short
30 Ohm Range
— +: 00.20335 ohm
= 00.20335 ohm
2
Short
300 Ohm Range
+ _000.2035 ohm
= 000.2035 ohm
3
Short
3 KOhm Range
+: 0.000207 Kohm
- 0.000207 Kohm
4
Short
30 KOhm Range
+ 00.00027 Kohm
- 00.00027 Kohm
5
Short
300 KOhm Range
+ 000.0010 Kohm
+ 000.0010 Kohm
6
Short
3 Mohm Range
+ -0.000014 Mohm
+ 0.000014 Mohm
7
Short
30 MOhm Range
+ 00.00083 Mohm
- 00.00083 Mohm
GAIN TEST
8
30 Ohm
30 Ohm Range
30.20560 ohm
29.79440 ohm
9
300 Ohm
300 Ohm Range
300.2200 ohm
299.7800 ohm
10
3 KOhm
3 KOhm Range
3.000357 Kohm
2.999643 Kohm
11
30 KOhm
30: KOhm Range
30.00177 Kohm
29.99823 Kohm
12
300 KOhm
300 KOhm Range
300.0160 Kohm
299.9840 Kohm
13.3 MOhm
3 MOhm Range
3.000209 Mohm
2.999791 Mohm
14
30 Mohm
30 MOhm Range
30.01283 Mohm
29.98717 Mohm

Advertisement

Table of Contents
loading

Table of Contents