Table 1-1. Specifications
FREQUENCY AND PERIOD (Cont'd)
SENSITIVITY:
10mv rms or 100UA rms sine-wave
TRIGGERING:
Triggers
and counts
on zero crossings.
MEASUREMENT
ACCURACY:
+(% of reading)
FREQUENCY
PERIOD
% OF
READING
VOHz
-
400Hz
100msec
-
2.5msec
0.05
400Hz
-
1.5MHz|
2.5msec
667nsec
0.01
MAXIMUM READING RATE:
2 readings/second
(Integration
time
of
1 PLC,
fast
settling
time,
delay
zero
and
math
OFF,
and
fixed
range)
PLUG-IN OPTION
44491A
Armature
Relay Multiplexer
Assembly
INPUT CHARACTERISTICS:
Eight two-wire armature relay channels and two current/actuator
channels.
MAX4MUM
VOLTAGE:
(Terminal
to
Terminal
or
Terminal
to
Chassis)
250 VOC or 250 VAC rms
MAXIMUM CURRENT: (per channel)
1 amp DC or AC rms
THERMAL OFFSET:
< 3 microvolts
RELAY
LIFE:
107
operations at maximum load (1.5 amps AC rms).
CLOSED CHANNEL RESISTANCE:
(end of relay life)
< 2 ohms
MAXIMUM SWITCHING AND MEASUREMENT
SPEED:
33
channels/second
ADJACENT CHANNEL CROSSTALK:
(Channels terminated
into 5U ohms)
+60
db
@ 100KHz

Advertisement

Table of Contents
loading