stimulus during the multimeter Self Test routine.
8-14,
AC to DC Converter
The AC to DC Converter is a computing true rms converter which converts AC input signals up to 10
volts full scale and between
20 Hz and
| MHz
to an equivalent
DC
voltage for measurement
by the
HP 3457A.
8-15.
Analog to Digital (A/D) Converter
The Analog to Digital Converter
uses a multi-slope conversion
process to convert
the analog measure-
ment data to digital information.
8-16. Voltage Reference
The Voltage Reference circuit, as the name
implies, provides a stable voltage to the Analog to Digital
Converter and the Ohms Current Source for measurement reference.
8-17.
A/D Controller
The A/D Controller controls the measurement
process,
Function and range information received from
the Main Controller is used to control the input switching
and A/D converter.
Measurement
values are
computed
by the A/D Controller and then transmitted
to the Main Controller
through
the isolation