SECTION III
OPERATION
3-1. INTRODUCTION
3-2. The information
contained
in this section describes manual
operation
of the HP 3457A
Multimeter,
The contents are intended to meet the needs of service personnel with emphasis on front panel operation,
For more detailed information on overall operation, refer to the HP 3457A Operating Manual.
3-3. GENERAL
INFORMATION
3-4. AC Power
3-5. Before connecting
power to the HP 3457A, be certain the line switches on the rear panel are set for
your power source and that the proper
fuse is installed. Information
for setting the line switches and for
fuse selection is located in Section II of this manual.
3-6. Power-On
3-7. The HP 3457A
automatically performs a power-on self test when it is switched on, This test takes ap-
proximately
1.5 seconds to complete and basically assures that the instrument
is capable of operating. The
test does not necessarily
indicate that measurements
will be accurate. The power-on
self test checks
the
master processor, slave processor and the communication
isolation circuitry (three of the ten tests perform-
ed by the complete self test). Upon satisfactory completion of the test, the instrument
will display its HP-IB
address for approximately five seconds.
The address is displayed as ADDRESS-dd
where "dd" is the decimal
address code of the instrument (the address is set to decimal 22 at the factory). If the test can not be com-
pleted, the instrument
will display
FAILED.
If this occurs,
refer to Section
VIII for troubleshooting
procedures,
3-8. Upon completion of the Power-On
Test, the instrument sets itself to predefined conditions (Power-On
state). The power-on state conditions are listed in Table 3-1.
3-9. OPERATION
3-10. The following paragraphs des
sic Operating procedures for the
HP 3457A
Multimeter,
This sec-
tion covers only front panel operating
procedures such as voltage, resistance, and current
measurements.
Also included
are manual
procedures covering operation
of the HP 44491A
and 44492A
Multiplexer
Assemblies. For detailed operating information, refer to the HP 3457A Operating Manual.
3-11. Voltage Measurements (Manual Operation)
3-12. Function Selection.
The measurement
function is selected by pressing the appropriate key. The volt-
age measurement functions available are DC Volts, AC Volts and DC Coupled AC Volts. The default func-
tion is DC Volts.
® To measure DC Voltage - Press the DCV function key.
© To measure AC Voltage - Press the ACV / ACDCV function key.
© To measure DC coupled AC voltage - Press the Blue SHIFT key then the ACV / ACDCV
function key.
3-1