Table 1-1. Specifications
TRUE RMS AC VOLTAGE (Cont*d)
90 DAY:
Teal
+ 5°C (ACDCV
Function)
After
1 hour warm-up
30 mV to 300 V ranges
FREQUENCY
* PERCENT
NUMBER OF COUNTS
OF
>= 1 PLC
0.1 PLC
+005 PLC
-0005 PLC
READING
|6 1/2 digits|5
1/2 digits|4
1/2 digits|3
1/2 digits
20Hz-45Hz
1.36
3600
364
41
45Wz-100Hz
0.17
3600
364
41
10OHz-20KHz
0.17
3600
364
4)
** 4OOHZ-20KHZ
0.44
3600
2810
285
20Kiz-100KHz]
0.66
4620
466
51
100KHz-300KHz|
3.16
11400
1144
119
300KH2-1MHz
10.16
69600
6960
701
90 DAY: Teal
+ 1°c CACDCV Function) After
1 hour
warm-up
300 V range
FREQUENCY
PERCENT
NUMBER OF COUNTS
OF
>= 1 PLC
0.1 PLC
-005 PLC
-0005 PLC
READING
|6 1/2 digits]5
1/2 digits|4
1/2 digits]3
1/2 digits
** 4OOHZ-20KHZ
0.5
3600
2810
285
33
20KHz-100KHz
1.16
6420
650
69
WW
* Specifications
apply within one week of AC auto-cal
(ACAL 2) for stable temperature conditions
(Teal
£5°C).
If AC auto-cal
is not used, add 0.6 to the percent of reading figures,
add 39000
to the 6 1/2 digit counts,
add 3900 to the 5 1/2 digit counts,
add 390 to the 4 1/2 digit
counts
and
39
to
the
3 1/2
digit
counts.
Using AC fast filter (ACBAND 400) for
frequencies above 400Hz.
For 1 year specifications,
add .1% to 24 hour
percent of reading figures,
add 2100 to 6 1/2
digit counts,
add 210 to the 5 1/2 digit counts,
add 21 to the 4 1/2 digit counts,
add 2.1 to
the
3 1/2
digit
counts.
TEMPERATURE
COEFFICIENT
- DC COUPLED:
300 mV, 30 V and 300 V rang
FREQUENCY
PERCENT
NUMBER OF COUNTS
READING
|6 1/2 digits|5
1/2 digits]4
1/2 digits|3
1/2 digits
20Hz-100KHz|
0.07
1000
100
10
VO1KHz-
MHz
0.08
1300
130
13
1-15

Advertisement