HP 3457A Service Manual page 130

Hide thumbs Also See for 3457A:
Table of Contents

Advertisement

Calibration
3457A
c. Use the following procedure to calibrate the AC Current function on the 30 mA, 300 mA and | A
ranges beginning with the 30 mA range.
S
1, Set the 3457A to the appropriate current range. (Use the { or | scroll key to select the proper range).
2. Set the current source to provide a full-scale reading on the HP 3457A range selected.
3. Access the Calibration Command.
(Press the Blue SHIFT key, then the OFFSET COMP / C configura-
tion key). The display will show CAL.
4. Enter the value of the input current (in amps).
(NOTE: If the security feature has not been disabled
(set to 0), it will be necessary to enter the input value, the delimiter ( , ) and the security code).
5, Press the ENT
key. The display will show
CALIBRATING
while the calibration
routine is
running.
d, Repeat steps "cl" through "c5" for each of the remaining ranges.
e. Set the Current Source for an output of 0 and disconnect it from the HP 3457A,
5-43. Frequency Calibration - Front Terminals
5-44. Equipment Required. A Frequency Generator capable of providing a single sine-wave signal between
100 Hz and | MHz with a frequency accuracy of +.003%) is required to calibrate the Frequency Function.
a, Set the 3457A to the Frequency function,
(Press the FREQ function key ).
b, Set the Frequency Source to a known frequency between
100 Hz and | MHz and connect it to the
a
HP 3457A front panel HI and LO input terminals.
c. Access the Calibration command.
(Press the Blue SHIFT key, then the OFFSET COMP / C configuration
key). The display will show CAL.
d. Enter the value of the input frequency (in hertz). (NOTE: If the security feature has not been disabled (set
to 0), it will be necessary to enter the input frequency, the delimiter ( , ) and the security code),
e. Press the ENT key. The display will show CALIBRATING while the calibration routine is running.
f. Disconnect the Frequency Source from the HP 3457A,
5-45. CALIBRATION PROCEDURES - REAR PANEL INPUTS
5-46. The following paragraphs
describe the calibration
process for the HP 3457A
rear panel
inputs.
Separate
procedures
are provided
for the standard
instrument
(Rear Input Terminals), and instruments
equipped with optional plug-in assemblies 44491A (General Purpose Relay Assembly) and 44492A (Reed
Relay Multiplex Assembly). Select the procedure which applies to your instrument.
5-47. DC Volts Offset Calibration - Rear Terminals (Standard Instrument)
5-48. Equipment Required. A low thermal short (copper wire) is required for this procedure.
a. Short the Q SENSE HI and LO and INPUT HI and LO rear input terminals together.
5-12