HP 3457A Service Manual page 82

Hide thumbs Also See for 3457A:
Table of Contents

Advertisement

3457A
Operational
Verification Tests
AC Current Operational Verification Tests
ATH ————_—
RECALL
STORE)
_ sao
MEAN
A SOEV
A_NSAMP | ERROR
©
C4 I
°
AC
[ ice |
" u g
| | ) | CURRENT
:
SOURCE
Figure 4-15. ACI Operational Verification Test Connections
4-78. AC Current Function - Gain Test
4-79, Equipment Required, An AC Current Source capable of providing a current of 300 mA
(+ .1%) at a
frequency of | KHz is required for the following test.
1, Set the AC Current source for an output of 0 mA and connect it to the HP 3457A front panel I
and LO input terminals.
2. Set the HP 3457A
to the 300 mA
range and the number of digits displayed (DIGITS DISP) to six.
Set the AC Current Source for an output of 300 mA at a frequency of 1 KHz.
3. Use the Test Record to record the 300 mA full-scale reading.
4, If the full-scale reading is beyond the limits specified on the Test Record (301.0300 mA - 298.9700
mA), the instrument should be calibrated or repaired. Refer to Section V for calibration procedures.
5. Reduce
the output of the AC Current
Source to 0 and disconnect
it from
the HP 3457A
input
terminals.
2-Wire Ohms Operational Verification Tests
4-80. 2-Wire Ohms Function - Offset Test
4-81. Equipment Required. A low thermal short (copper wire) is required for this procedure.
1, Connect a low thermal short across the HP 3457A front panel HI and LO input terminals.
2. Set the HP 3457A to the 2-Wire Ohms Function, 3 Kohm range and the number of digits displayed
(DIGITS DISP) to six.
3. Use the Test Record to record the offset readings for the 3 Kohm, 30 Kohm and 300 Kohm ranges.
4-27