HP 3457A Service Manual page 53

Hide thumbs Also See for 3457A:
Table of Contents

Advertisement

3457A
Operation
© Set the Scan
Advance
to Auto - Press the SCAN
ADV / O configuration
key. Use the f / AUTO
or
| / HOLD scroll key to display AUTO:2 and press the ENT / LAST ENTRY key.
© Enable the Reading Memory
- Press the CONFIG / R configuration key then the ENT / LAST
ENTRY
key.
© Allocate Reading Memory
space (each reading requires 4 bytes of memory) - Press the Blue SHIFT key
then the TERM / M configuration key. Use the | / HOLD scroll key to display MSIZE. Enter the number
40 (10 readings times 4 bytes) and press the ENT / LAST ENTRY key.
© Select the channels to be measured - Press the Blue SHIFT key then the RECALL / S configuration key.
Use 1 / HOLD
scroll key to display SLIST. Enter the number of each channel
you wish to measure
fol-
lowed by a comma (,) to separate them (example: 0,1,2,3,4,5,6,7,8,9). Press the ENT / LAST ENTRY
key, (As
many as 16 single digit channel numbers can be entered at one time. The number of readings (NRDGS)
should be set to match the number of channels entered).
© Make the measurement - Press the SINGLE / F configuration key to start the measurement sequence.
The instrument will read each channel listed and store the readings in memory.
© Recall the readings - Press the Reading Memory
RECALL / S key then the ENT / LAST
ENTRY
key.
The reading displayed
is the last reading taken. The readings are shifted
into memory
so that the first
reading taken is in the highest number memory
location. The last reading taken is in memory
location
1.
Use the t / AUTO and J / HOLD scroll keys to view the measurement
readings. Use the + and > display
keys to read the total display contents.
©
To
repeat
the same
set
of
measurements
-
Press
the
CONFIG/R
configuration
key
then
the
ENT / LAST ENTRY
key to clear the reading memory. Press the SINGLE / F configuration key to start
the new reading sequence.

Advertisement