Table 1-1. Specifications
TRUE RMS AC CURRENT
(Cont'd)
24
HOUR:
Teal
+
1 A range. (1 amp maximum input)
FREQUENCY
PERCENT
OF
READING
>= 1 PLC
6 1/2
digits
NUMBER O}
0.1 PLC
5 1/2
digits
F COUNTS
-005 PLC
4 1/2 digits
General Information
1°C (ACDCI
Function) After 2 hour warm-up
Accuracy
relative to calibration std
-0005 PLC
3 1/2 digits
1.6
0.45
0.35
0.7
16000
16000
16000
16000
20Hz
46Hz
1VO1HZ
**
400HZ
90 DAY:
Tcal +
30
mA
and
300
mA
ranges.
* PERCENT
OF
READING
FREQUENCY
>= 1 PLC
6 1/2 digits
1600
1600
1600
3750
5°C (ACDCI
Function) After 1 hour warm-up
NUMBER O}
0.1
PLC
5 1/2
digits
165
165
165
375
F COUNTS
005
PLC
4 1/2
digits
20
20
20
42
-0005
PLC
3 1/2 digits
16000
16000
16000
16000
17500
20Hz-45Hz
46Hz- 100Hz
i
101Hz-20KHz
3
**
4OOHZ-20KHz
65
21KHz-
100KHz
95
+55
90 vaY: Teal +
1 A range. (1 amp maximum input)
* PERCENT
OF
READING
FREQUENCY
>= 1 PLC
6 1/2
digits
1600
1600
1600
3750
1750
1°c (ACDCI] Function) After 1 hour warm-up
NUMBER
O|
0.1 PLC
5 1/2 digits
165
165
165
375
180
F COUNTS
-005
PLC
4 1/2 digits
20
20
20
42
22
+0005 PLC
3 1/2
digits
20
20
20
42
1600
1600
1600
3750
165
165
165
375
1.65
0.5
0.4
0.75
16000
16000
16000
16000
20Hz-45Hz
46Hz-100Hz
1O1HZ-20KHz
**
4OOHZ-20KHz
For 1 year specifications, add .08% to the 90 day Percent of Reading figures.
**
Using
AC
fast
filter
(ACBAND
400)
for
frequencies
above
400Hz.
1-19