For installations including a Residual-Current Circuit Breaker (RCCB):
• Only use an industrial RCCB with, for example, a 300 mA rating.
• Install in accordance with the national requirements for industrial electrical installations.
6 6 . . 2 2 . . 2 2 P P o o w w e e r r C C o o n n s s u u m m p p t t i i o o n n
M M a a x x i i m m u u m m p p o o w w e e r r c c o o n n s s u u m m p p t t i i o o n n (melting with full
beam current)
I I d d l l e e ( ( m m a a c c h h i i n n e e o o n n , , n n o o v v a a c c u u u u m m ) )
A A v v e e r r a a g g e e p p o o w w e e r r c c o o n n s s u u m m p p t t i i o o n n d d u u r r i i n n g g a a b b u u i i l l d d
N N o o t t e e
Due to phase shift in the system, the apparent power value is significantly higher than the active
power. Individual phases demand up to 20 A for an active system. Initial peaks (duration of milli-
seconds) exceeding 32 A exist when turning the system on.
In power supply systems with a Residual-Current Device (RCDs), the RDC should have a break-
ing current of 300 mA.
01
IM-710429
en-GB
5.9 kW (12.6 kVA)
1.5 kW (2.0 kVA)
3.9 kW (8.0 kVA)
35