UPS Run Time Estimates
Grounding
Revision C
Installation: Installation Requirements
&$87,21
Connect the UPS to the ApexPro Telemetry Server(s) only. Do
NOT connect printers or other devices to the UPS, as such devices
may shorten estimated run-times. If AC line power is not restored
before UPS run-time is exceeded, the ApexPro Telemetry Server
shuts down and patients will not be monitored.
UPS run-times are affected by:
Battery age,
Ambient temperature,
Site specific UPS usage patterns, and
Load characteristics.
Your actual run times may be different.
UPS Description
Power Supply UPS 600VA 120 Volts
Power Supply UPS 600VA 220 Volts
Power Supply UPS 750VA 120 Volts
Power Supply UPS 750VA 220 Volts
Power Supply UPS 2000VA 120 Volts
Refer to
"Optional Components"
To ensure safety and for proper operation, the ApexPro Telemetry Server must be
properly grounded in accordance with all applicable IEC and UL specifications,
applicable building codes, and local electrical inspection regulations. For
ApexPro Telemetry Server
2001989-178
Runtime Estimates
1 ApexPro Telemetry Server for approximately
40 minutes
2 ApexPro Telemetry Server for approximately
15 minutes
1 ApexPro Telemetry Server for approximately
42 minutes
2 ApexPro Telemetry Server for approximately
14 minutes
1 ApexPro Telemetry Server for approximately
48 minutes
3 ApexPro Telemetry Server for approximately
11 minutes
1 ApexPro Telemetry Server for approximately
41 minutes
3 ApexPro Telemetry Server for approximately
10 minutes
1 ApexPro Telemetry Server for approximately
1 hour, 22 minutes
5 ApexPro Telemetry Server for approximately
9 minutes
on page 6-3 for UPS part number information.