Symptom
POST error after scanning
IDE drives...
Message: "No drives
attached to FastTrack
controller..." then fails to
"Disk boot failure".
Inability to access SNMP
while server is shut down.
Inability to access SNMP
while server is either shut
down or running.
Inability to read
temperature via SNMP.
System overheating.
Inability to use CDROM
drive.
5-12
Troubleshooting: Component Troubleshooting
Possible Reason
Loss of both hard drives.
Discharged SNMP card
battery.
Dead SNMP card battery.
incorrect SNMP card
battery installation
(reversed polarity).
SNMP card failure.
Incorrectly connected
SNMP cable.
Misconfigured IP address.
Temperature sensor failure.
Dust or debris in fans or
filters.
System fan failure.
Failure of CDROM drive.
Loose or disconnected
CDROM drive cable.
ApexPro Telemetry Server
2001989-178
Solution/Course of Action
Replace both hard drives.
AC power outage will
discharge SNMP battery.
Plug ApexPro Telemetry
Server into a working outlet
and recharge battery
Replace SNMP card
battery.
Re-install SNMP card
battery.
Replace SNMP card.
Verify and/or re-install
cable.
Reconfigure IP address.
Replace temperature
sensor.
Clean system.
Replace fans.
Replace CD drive.
Reconnect drive cable.
FRU Procedure
See
"Hard Drive"
on
page 6-30.
N/A
See
"SNMP Battery"
on
page 6-24.
See
"SNMP Battery"
on
page 6-24.
See
"SNMP Boards"
on
page 6-18.
N/A
N/A
See
"Temperature
Sensor"
on page 6-52.
See
"Fan Filter"
on
page 6-7.
See
"Drive Filter"
on
page 6-9.
See
"Chassis Fan"
on
page 6-38.
See
"CD ROM Drive"
on
page 6-34.
N/A
Revision C