Stop/Start/Restart ApexPro Telemetry Server

Revision C
Equipment Overview: Service Utilities
:$51,1*
LOSS OF MONITORING — Before stopping or restarting the
ApexPro Telemetry Server software, or shutting down or
restarting the ApexPro Telemetry Server computer, notify the
site's biomedical department so patient communications can be
established with an alternative central station (the Bed Duplicator
application can be used to temporarily transfer patients to an
alternate ApexPro tower. See
page 4-4). Shut down of the ApexPro Telemetry Server causes
disruption of monitoring function at the CIC Pro.
1.
Log in to the Webmin application according to
21.
2.
Click the Configuration tab.
3.
Click the Run link.
4.
Select one of the following:
To stop the ApexPro Telemetry Server software, click Stop, next to
ApexPro Commands.
To restart the ApexPro Telemetry Server software, click Restart, next to
ApexPro Commands.
To shutdown the ApexPro Telemetry Server computer, click Shutdown,
next to Operating System Commands.
To restart the ApexPro Telemetry Server computer, click Restart, next to
Operating System Commands.