1.5.4 Selecting Mass Flow Units

1. Scroll to the desired Mass Flow units for the flow rate display and press [Enter]. The available units for this prompt
are determined by the selection made at System Units (see Table 2).
2. Scroll to the desired Mass Flow Time units for the mass flow rate display and press [Enter].
3. Scroll to the desired number of Mdot Decimal Digits (digits to the right of the decimal point in the mass flow rate
display) and press [Enter].
4. Scroll to the desired Mass (Totalizer) units for the totalized mass flow rate display and press [Enter]. The available
units for this prompt are determined by the selection made at the System Units prompt.
5. Scroll to the desired number of Mass Decimal Digits (digits to the right of the decimal point in the totalized mass
flow rate display) and press [Enter].
Procedure Options
After completing the steps on the previous page, the XGS868i returns to the Channel PROGRAM window. Do one of
the following:
To continue entering "quick startup" data, continue to Step 1 in Entering Transducer and Pipe Parameters on the
next page.
To continue regular programming, refer to Appendix A, Menu Maps, to navigate to the desired menu.
To leave the Keypad Program, press [Escape] twice.
DigitalFlow™ XGS868i Programming Manual
Table 2: Available Mass Flow Units
English
LB = Pounds
KLB = Thousands of LB
MMLB = Millions of LB
TONS = Tons (2,000 LB)
Metric
KG = Kilograms
TONN = Metric Tons (1,000 KG)
Chapter 1. Programming Site Data
7