GE DigitalFlow XGS868i Programming Manual page 41

Panametrics steam ultrasonic mass flow transmitter
Table of Contents

Advertisement

1.9.2f Alarm Relays (cont.)
Selecting the Alarm Type (cont.)
6. Do one of the following:
high or low - proceed to Step 7 below.
fault - proceed to Step 9 below.
7. Scroll to the desired measurement parameter and press [Enter]. (See Table 9 on page 26 for a description of the
available options.)
Note: The measurement units that appear in these prompts are those selected in the
this section.
8. Enter a value for the Trigger Point of the alarm and press [Enter]. Go to Procedure Options below.
9. Scroll to the type of Error (flow, non-flow or both) that will trigger the fault alarm and press [Enter].
Procedure Options
After completing the previous steps, the XGS868i returns to the I/O window. Do one of the following:
To continue programming, refer to Appendix A, Menu Maps, to navigate to the desired menu.
To leave the Keypad Program, press [Escape] three times.
DigitalFlow™ XGS868i Programming Manual
Chapter 1. Programming Site Data
Global-System
window earlier in
31