Pipe Data - GE DigitalFlow XGS868i Programming Manual

Panametrics steam ultrasonic mass flow transmitter
Table of Contents

Advertisement

1.6.2 Pipe Data

If either a standard or a special transducer is being used, the programming sequence should be rejoined at this point.
5. To select the appropriate Pipe OD Unit type from the list shown in Table 3 below, scroll to the right side of the
screen, and use the up and down arrow keys to step through the list. Press [Enter]. Then use the arrow keys to enter
the known pipe outside diameter or circumference on the left side and press [Enter].
Obtain the required information by measuring either the pipe outside diameter (OD) or circumference at the
transducer installation site. The data may also be obtained from standard pipe size tables found in Sound Speeds
and Pipe Size Data (914-004).
6. Use the arrow keys to enter the known Pipe Wall Thickness (in inches or mm) and press [Enter]. If the pipe wall
thickness is not available, look up the value in a table of standard pipe size data which can be found in the Sound
Speeds and Pipe Size Data manual (914-004)
IMPORTANT:
Because the units cannot be independently chosen for this parameter, the value must be entered in the
same units used for the pipe OD/Circumference.
1.6.2a
Path and Axial Lengths
7. To enter the Path Length:
a. Use the [] arrow key to highlight the path length unit type at the right of the screen.Then use the [] and []
arrow keys to scroll to the desired unit type.
b. Use the [] arrow key to return to the numeric entry on the left, and enter the path length of the ultrasonic signal.
Press [Enter].
DigitalFlow™ XGS868i Programming Manual
Table 3: Available Pipe Size Units
English
inches
feet
in/PI = pipe circumference
in inches
ft/PI= pipe circumference
in feet
Metric
mm = millimeters
m = meters
mm/PI = pipe circumference
in millimeters
m/PI = pipe circumference
in meters
Chapter 1. Programming Site Data
9

Advertisement

Table of Contents
loading

Table of Contents