Navigator 500
Sodium analyzer

8.4 Manual regeneration

Referring to Fig. 8.3:
1. Carefully unscrew plastic sleeve
B
sodium electrode
in place).
2. Withdraw sodium electrode
C
flowcell
(it is not necessary to detach the electrode
D
lead
).
E
3. Remove O-ring
.
4. Prepare two plastic beakers, one containing about 50 ml
of etch solution, the other about 200 ml high purity water.
5. Dip the electrode in the etch solution for 60 (±5) seconds
then rinse in high purity water.
Caution. Do not to exceed the etch time or the
electrode's performance may be permanently
degraded.
6. Dispose of the etch solution safely by diluting with plenty
of water (use fresh etch solution each time).
E
7. Refit O-ring
, sodium electrode
A
sleeve
finger-tight (do not over-tighten).
8. Run the wet-section for 1 to 2 hours on sodium sample
before attempting a calibration.
Fig. 8.3 Manual regeneration of sodium electrode
OI/ASO550–EN Rev. G
A
a few turns (with
B
from the sleeve and
B
and screw plastic

8.5 Shut-down procedures

Warning. Observe relevant safety procedures when
handling reagent solutions.
1. Close the sample valve upstream of the wet-section.
2. Remove the reagent container (see Fig. 2.1, page 6 for
bottle locations) and store or dispose of the solution safely.
If disposing, rinse the container thoroughly.
3. Fill the regeneration solution container (if fitted) with high
purity water and energize the regeneration valve via the
Man. Control Valve menu (see Section 7.1.2, page 47) for
5 minutes.
4. Fill the calibration solution containers with high purity
water and perform a two-point calibration to flush the
wet-section.
5. Remove the electrodes and follow storage procedure in
Section 8.5.1.
6. Use a syringe to flush all tubing with high purity water. This
removes any particulate deposits.
7. Switch off the mains supply to the transmitter.
Note. For multiple wet-section installations, do not
switch off the mains supply to the transmitter if
operation of the remaining wet-sections needs to be
continued.
8 Maintenance