4. If the Calibrate page is not displayed, use the
keys to scroll to the Calibrate page:
Menu
Calibrate
Exit
5. Press the
key (below the Select prompt).
–
If one or more single-stream wet-sections are
connected, the Calibrate / Sodium 1 (2, 3, 4) page is
displayed:
Calibrate
Sodium 1
Back
–
If a multi-stream wet-section is connected, the
Calibrate / Sodium Three Stream page is displayed:
Calibrate
Sodium Three Stream
Back
6. Press the
key (below the Select prompt).
The following page is displayed with the Sensor
Calibration menu selected:
Sensor 1
Sensor Calibration
Scheduled Calibration
Calibration Setup
Restore Defaults
Back
62
/
Select
Select
Select
Select
7. Press the
key (below the Select prompt).
The Sensor Calibration page is displayed:
Sensor Calibration
Low Calibration
High Calibration
Two-Point Cal
Regeneration Cycle
Process Calibration
Back
Press the
key to highlight the Regeneration Cycle
menu.
8. Press the
key (below the Select prompt).
The Regeneration Cycle page is displayed:
Regeneration Cycle
Start Procedure
No
Press the
key (below the Yes prompt) to start the
regeneration procedure.
8 Maintenance
Select
Yes