6.8.5 Log files

The Alarm Event, Calibration, Diagnostic and Audit logs are
archived into the same file. The filenames are formatted as
follows:
Event logs: <ddmmyy><hhmmss><instrument tag>.AOO

6.8.6 Daylight saving

Files containing data generated during the daylight saving period
have '~DS' appended to the filename.
Start of daylight saving period
A daily file is started at 00:00:00 on 30
30Mar13ASO550.D00
Summertime starts at 2:00am on 30
changes automatically to 3:00am.
The existing file is closed and a new file is created filename:
30Mar13ASO550~DS.D00
The file '30Mar13ASO550.D00' contains data generated from
00:00:00 to 01:59:59.
The file '30Mar13ASO550~DS.D00' contains data generated
from 03:00:00.
End of daylight saving period
A daily file is started at 00:00:00 on 26
26Oct13ASO550~DS.D00
Summertime ends at 3:00am on 26
clock changes automatically to 2:00am.
The existing file is closed and a new file is created filename:
26Oct13ASO550.D00
The file '26Oct13ASO550~DS.D00' contains data generated
from 00:00:00 to 02:59:59.
The file '26Oct13ASO550.D00' contains data generated from
02:00:00.
Note. Daily files start at 00:00:00.
38
th
March 2013 filename:
th
March 2013 and the clock
th
October 2013 filename:
th
October 2013 and the
6 Operation overview – transmitter