Electrical Connections - Transmitter - ABB Navigator 500 Series Operating Instructions Manual

Sodium analyzer
Hide thumbs Also See for Navigator 500 Series:
Table of Contents

Advertisement

Navigator 500
Sodium analyzer
3.5 Electrical connections – transmitter
Warning.
If the transmitter is used in a manner not specified by the Company, the protection provided by the equipment may be
impaired.
Remove all power from supply, relay, any powered control circuits and high common mode voltages before accessing or
making any connections. Use cable appropriate for the load currents: 3-core cable rated 3 A and 75 °C (167 °F) minimum,
and voltage: 100 / 240 V that conform to either IEC 60227 or IEC 60245, or to the National Electrical Code (NEC) for the US,
or the Canadian Electrical Code for Canada. The terminals accept cables AWG 26 to 16 (0.14 to 1.5 mm
Ensure the correct fuses are fitted – Fig. 3.14, page 18 for fuse details.
Replacement of the internal battery must be carried out by an approved technician only.
The transmitter conforms to Installation Category II of IEC 61010.
All connections to secondary circuits must have insulation to required local safety standards. After installation, there must be
no access to live parts, for example, terminals. Use screened cable for signal inputs and relay connections. Route signal
leads and power cables separately, preferably in an earthed (grounded) flexible metal conduit.
All equipment connected to the transmitter's terminals must comply with local safety standards (IEC 60950, EN61010-1).
The ethernet and bus interface connectors must only be connected to SELV circuits.
USA and Canada Only
The supplied cable glands are provided for the connection of signal input and ethernet communication wiring ONLY.
The supplied cable glands and use of cable / flexible cord for connection of the mains power source to the mains input and
relay contact output terminals is not permitted in the USA or Canada.
For connection to mains (the mains input and relay contact outputs), use only suitably rated field wiring insulated copper
conductors rated min. 300 V, 16 AWG, 90C. Route wires through suitably rated flexible conduits and fittings.
16
3 Installation
2
).
OI/ASO550–EN Rev. G

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Navigator 550Navigator 540

Table of Contents