Specification - Analyzer - ABB Navigator 500 Series Operating Instructions Manual

Sodium analyzer
Hide thumbs Also See for Navigator 500 Series:
Table of Contents

Advertisement

Navigator 500
Sodium analyzer
9 Specification – analyzer
Operation
Measuring range
0.01 to 10,000 ppb
Units of measure
ppb, µg/l, µg/kg
Accuracy
±5 % of reading or ±0.1 ppb, whichever is the greater (applies only
when sample is within ±5 °C of calibration temperature)
Repeatability
±5 % of reading or ±0.1 ppb, whichever is the greater (applies within
a ±5 °C [±9 °F] variation)
Response time
1 to 100 ppb <4 mins;
100 to 1 ppb <4 mins for a 90 % step change
Multi-stream measurement
Sample time programmable between 10 and 60 min.
Resolution
0.001 ppb
Temperature compensation
5 to 55 °C (41 to 131 °F) automatic using a Pt1000
AutoCal frequency
Programmable from 1 to 7 days or 1 to 8 weeks
Sample temperature
5 to 55 °C (41 to 131 °F)
Sample pressure
1.5 bar gauge (21.7 psi) maximum
Sample flow rate
100 to 400 ml/min
Sample connections
1
/
ID flexible tubing to barbed connector
4
OI/ASO550–EN Rev. G
Environmental data
Ambient operating temperature:
0 to 55 ºC (32 to 131 ºF)
Ambient operating humidity:
Up to 95 % RH non-condensing
Storage temperature:
– 20 to 70 °C (–4 to 158 °F) without sensor
0 to 55 °C (41 to 131 °F) with sensor
Approvals, certification and safety
CE mark
cULus
General safety
EN61010-1
Pollution degree 2
Insulation class 1
EMC
Emissions & immunity
Meets requirements of IEC61326 for an industrial environment and
domestic emissions
Maintenance
Periodic calibration:
User-defined
9 Specification – analyzer
DS/ASO550–EN Rev. E
69

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Navigator 550Navigator 540

Table of Contents