7.1.2 Sensor Setup

Menu
Sensor Setup
Select
Exit
Menu
Sodium 1 – 4
Tag
Stream 1 (3) Tag
Units
Range High
Stream 1 (3)
Range High
Range Low
Stream 1 (3)
Range Low
Filter Type
Filter Time
Wetting Routine
46
Used to set the wet-section tag, measurement units, operational and stream ranges and filter type
and enable / disable the Flow Meter (optional) and Manual Valve Control functions.
Comment
The identity of the sodium wet-section being set up (if a sodium
multi-steam wet-section, Sodium only is displayed).
Note: Single-stream wet-sections only.
Sets the (colored) 16-character alphanumeric wet-section tag
displayed on Operator Pages.
Note: Multi-stream wet-sections only.
Sets the 16-character alphanumeric tag for each selected stream
displayed on Operator Pages.
Selects the measurement units:
ppb / µg/l / µg/kg.
Note. Single-stream wet-sections only.
Sets the range span in Chart and Bargraph views.
Note. Multi-stream wet-sections only.
Sets the range span in Chart and Bargraph views.
Note. Single-stream wet-sections.
Sets the range zero in Chart and Bargraph views.
Note. Multi-stream wet-sections only.
Sets the range zero in Chart and Bargraph views.
Selects the filter type:
Off / Min / Max / Average / Rolling Point.
Sets the filter (input) time (5 to 100 seconds).
Enables / disables the wetting routine. The wetting routine is initiated
when a No Samples situation is detected. It energizes the high
calibration solution valve for 1 minute every 24 hours.
The wetting routine is terminated if sample returns.
7 Menu descriptions
Default
TAG 1
ppb
Off
5 seconds