Advertisement

5
Digital outputs
There are four digital outputs (electrically isolated passive output circuits) in the
connection box (TB6-terminals, details see figure).
The output at TB6-5 is a breaker (NC, "normally closed"), the other outputs are
makers (NO, "normally open"). Output DO1 is an NC contact (terminal TB6-1) and
output D02 is a NO contact (terminal TB6-3).
The maximum load per channel is 28.8 V DC/120 mA. The maximum pulse rate is
25 Hz.
Cable type refer to section
the maximum length is 250m ≙ 820ft. The figure also shows the appropriate ter-
minals in the connection box. The wiring in the drawing is only designed to il-
lustrate the screening using an example.
These outputs act like a potential-free contact to a common point. This
common point is usually used as common supply (usually +24 V). If con-
figured that way and a channel is switched ON, the corresponding pin
switches to +24 V. In addition to the wiring, various enSuite settings are
required before this interface can be used. Refer to section
I/O (IP address network- and I/O-settings)
outputs can be observed and manually set by setting these outputs into
"Test mode" on the display.
GasLab Q2
GasLab Q2 after delivery and at the place of use
5.6.2 Cables (power supply / communication)
➪  
Fig. 5-10: Example connection of digital outputs
User Manual (N )
8.2.1 SFB
➪  
(p. 190).The state of these
89 of 315
(p. 76)