8
8.2.6
Fiscal parameters and optional using of
approval file
For devices, which are intended to be used in legal metrology, the approval speci-
fies which parameters are legally relevant. The editing of such parameters must be
specifically protected or logged.
There are two different types of access rights for legally relevant para-
meters:
Can only be changed if the sealable security switch is open
Can also be changed if the security switch is closed, provided the
fiscal audit trail is not yet full. The change is logged in the fiscal audit
trail
If the device is operated in accordance with a valid approval, then the following
conditions need to be met in the parameterization:
The access rights for legally relevant parameters must be set in line with the
l
approval.
The software versions of the fiscal software components must comply with the
l
approval.
In order to ensure compliance with these conditions, please proceed as follows:
1. Select the approval to be applied in the Configuration tab, using the drop-down
list. All parameters and parameter branches whose official access rights (i.e.
or
) deviate from the selected approval are now labeled in bold orange. A blue
label indicates that official access rights have been newly added.
2. Now click on Make compliant. By doing this, the access rights for legally
relevant parameters are set in line with the selected approval. The orange la-
beling is replaced by a blue one as the changes have not yet been saved.
3. Check the listing of AFBs in the Configuration tab. If an AFB is still labeled in
bold orange here, then this means that a software version was selected here
which does not comply with the approval.
Select the correct version which is in line with the approval. (Versions in line with
the approval are listed in black, and versions which are not in line with the
approval are listed in orange. Versions that are not compatible with the current
Basic System appear in red.)
GasLab Q2
Configuration and analysis software enSuite
Fig. 8-30: Configuration – selecting the approval file
User Manual (N )
219 of 315