Honeywell enCore Series Manual page 265

Gas quality analyzer
Hide thumbs Also See for enCore Series:
Table of Contents

Advertisement

10
Gas leak detector for flammable gases, calibrated with status or value display
l
Steps
1. Apply gas and pressure to all inlet gas lines of the measuring device so that a
gas volume is present in the lines.
2. Use a gas leak detector to check all parts of the installation up to the instrument
for leaks.
3. If gas leak detector shows leaks, Immediately shut off the affected parts of the
gas line system and depressurize them.
Otherwise the check is finished.
Check the tightness downstream device
Prerequisite(s)
The device is in operation and gas is flowing through it.
l
Following tools and materials are available:
Gas leak detector for flammable gases, calibrated with status or value display
l
Steps
1. Use a suitable gas leak detector to check all parts of the installation behind the
device (exhaust gas lines) for leaks.
2. If gas leak detector shows leaks, take the device out of operation immediately
and shut off the affected parts of the gas pipe system.
leaks
(p. 265). Otherwise the check is finished.
Procedure in case of leaks
Steps
1. Immediately switch off the affected parts, depressurize and de-energize them.
2. Do not use the device elimination of leaks.
3. Eliminate / seal the leak!
4. Check the effectiveness of your measures! !
Repeat
10.3.3 Check the tightness of the installation
➪  
GasLab Q2
Maintenance, cleaning and repair
Procedure in case of leaks
➪  
User Manual (N )
(p. 265).
Procedure in case of
➪  
(p. 264)
265 of 315