Honeywell enCore Series Manual page 80

Gas quality analyzer
Hide thumbs Also See for enCore Series:
Table of Contents

Advertisement

5
The required accessories listed below are supplied in the recommended or ordered
version.
*    High number of cores for ATEX and IECEx only; low number of cores for all other certificates
**    To comply with the specified explosion protection, the cable gland must be sealed as specified by
the cable gland manufacturer
Alternativ screw-in parts for use in Japan
EX CABLE GLAND MODEL SFLU 10-M
EX CABLE GLAND MODEL SFLB/T 10-M
EX BLANKING ELEMENT PLUG-M10/M25/M32
Furthermore, many variations are possible, such as the use of tubes and individual
wires or conduit stop boxes if they are sealed or potted.
Under FM conditions these must be sealed within 18" (45cm). For ATEX and IECEx
this distance is 1xD from the enclosure as for a normal cable gland. Contact Honey-
well if you wish to deviate from the standard equipment.
If cable glands and dummy plugs are not pre- assembled on the device, the
following applies with regard to assembly and replacement:
Treat the thread of the screw-in parts with threadlocker! For this purpose, apply
l
Loctite 243, Loctite 511 or equivalent threadlocker to the thread flanks in the
last third of the internal thread.
GasLab Q2
GasLab Q2 after delivery and at the place of use
User Manual (N )
80 of 315

Advertisement

Table of Contents
loading

This manual is also suitable for:

Gaslab q2

Table of Contents