Electrical Requirements

NOTE:
GE HEALTHCARE requires a dedicated mains power line and Ground for the proper operation
of its Ultrasound equipment. This dedicated power line shall originate at the last distribution
panel before the system.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full
size Ground wire from the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral:
The dedicated line shall consist of one phase (two lines), not shared with any other circuit, and a full
size Ground wire from the distribution panel to the Ultrasound outlet.
NOTE:
Please note that image artifacts can occur, if at any time within the facility, the Ground from the
main facility's incoming power source to the Ultrasound unit is only a conduit.
2-2-5-1

Vivid S60N/Vivid S70N Power Requirements

Electrical specifications for the Vivid S60N/Vivid S70N system are as follows:
Table 2-2
Input Voltage
100V AC to 240V AC
2-2-5-2

Inrush Current

Inrush current is not a factor for consideration, due to the inrush current limiting properties of the power
supplies.
Voltage
100 V
240 V
2-2-5-3

Site Power Outlets

A dedicated AC power outlet must be within reach of the unit without requiring the use of extension
cords. Other outlets adequate for the external peripherals, medical and test equipment required to
support this unit must also be present and located within 1 m (3.2 ft) of the unit. Electrical installation
must meet all current local, state, and national electrical codes.
2-2-5-4

Mains Power Plug

The Vivid S60N/Vivid S70N portable ultrasound scanner is supplied with an AC power cable, as
standard. In the event that the unit arrives without a power cable, or a power cable fitted with an
incorrect plug, contact your GE dealer. When necessary, the installation engineer will supply the
appropriate power plug to meet the applicable local regulations.
1
EVISION
Electrical Requirements
Tolerances
Inrush Current
(Console Only)
4.5 A
2.3 A
Chapter 2 - Site Preparations
V
IVID
Op. Current
±10%
500VA
S60N/V
S70N
B
IVID
ASIC
Frequency
50-60 Hz
S
M
ERVICE
ANUAL
2-3