Probe Shelf Replacement Procedure
8-6-11-1
Tools
Use the appropriate screwdriver as indicated in the Probe Shelf Replacement procedure.
FRU Part #
Refer to
8-6-11-2
Time Required
45 min
8-6-11-3
Preparation
Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in
8-6-11-4
Probe Shelf Removal Procedure
1) Remove the Keyboard Assembly as described in the
Procedure on page 8 -
2) Working from below, and while carefully supporting the Probe Shelf Holder, remove the 4 retaining
screws - two on either side - as seen in
Retaining Screw
Figure 8-201 Probe Shelf - Viewed from Below indicating Retaining Screws
3.) Remove the Probe Shelf.
8-6-11-5
Probe Shelf Installation Procedure
1.) Return the Probe Shelf to its location on the Keyboard Interface Assembly and secure it with the
4 retaining screws (previously removed).
2.) Install the Keyboard Assembly as described in the
Procedure on page 8 -
3) Lower the console to the maximum down position.
4.) Turn ON power to the system.
8-166
Section 8-6 - Mechanical Platform Components - Replacement Procedures
1
EVISION
Table 9-7
.
on page 9-6
44.
47.
V
S60N/V
IVID
Operator Panel Keyboard Assembly Removal
Figure
8-201.
Operator Panel Keyboard Assembly Installation
S70N
B
S
IVID
ASIC
ERVICE
Power Shut Down on page 4 -