  • Page 1 GE Healthcare Technical Publications 5128706–1–100 Revision 1 SENOGRAPHE DMR+ Pre–Installation Manual do not duplicate Copyright 2006 by General Electric Co.
  • Page 2 Geräte bedienen oder sich selbst und andere Personen deren Bestrahlung aussetzen können. Vor Inbetriebnahme dieses Apparats sollte sich das qualifizierte und befugte Bedienungspersonal mit den geltenden Kriterien für den ge- fahrlosen Strahleneinsatz durch sorgfältiges Studium des Hefts Nr. 26 der Internationalen Kommission für Strahlenschutz (ICRP) vertraut...
  • Page 3: Table Of Contents

    GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 TABLE OF CONTENTS CHAPTER TITLE PAGE REVISION HISTORY ............
  • Page 4 GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 TABLE OF CONTENTS (CONT.) CHAPTER TITLE PAGE 4–5 Use Limitation ............
  • Page 5 GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 INSTRUCTIONS CONCERNING SAFETY WARNING X–RAY EQUIPMENT IS DANGEROUS TO BOTH PATIENT AND OPERATOR UNLESS ESTABLISHED SAFE EXPOSURE FACTORS ARE STRICTLY OBSERVED Though this apparatus is built to the highest standards of electrical and mechanical safety, the useful x–ray beam becomes a source of danger in the hands of the...
  • Page 6 GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 Blank page.
  • Page 7: Revision History

    GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 REVISION HISTORY DATE REFERENCE REASON FOR CHANGE New release based on document 2270822–100. November 29, 2004 5128706–100 rev. 1 Added information on compliance with IEC 601–1–2 Edition 2 New release based on document 5128706–100.
  • Page 8 GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 Blank page.
  • Page 9: Introduction

    GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 CHAPTER 1 – INTRODUCTION 1–1...
  • Page 10 GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 ILLUSTRATION 1–1 SENOGRAPHE BASIC CONFIGURATION 1–2...
  • Page 11: Purpose And Scope Of This Manual

    GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 SECTION 1 PURPOSE AND SCOPE OF THIS MANUAL This planning direction provides pre–installation data for the SENOGRAPHE DMR. It considers only product related pre–installation. SECTION 2 RESPONSIBILITY OF PURCHASER SECTION 3 PRODUCT IDENTIFICATION Components of the SENOGRAPHE DMR covered by this pre–installation manual comprise:...
  • Page 12 GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 OPTIONAL EQUIPMENT Stereotix 2. Accessory storage device. 24 x 30 Bucky + COMPRESSION PADDLE. Hydraulic chair. Rad shield 2 or 3. DATAFLASH. 1–4...
  • Page 13 GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 TABLE 1-1 PRODUCT STRUCTURE PRODUCT COMPLIES WITH THE FOLLOWING REGULATIONS: IEC601.2.7; UL187; CSA No. 22.114; CRF21 Part 1020.30; and NFC74100. IEC. 801.2; EN 61000–4–2 SENOGRAPHE DMR V2+ WITH XRAY PROTECTIVE SHIELD ACCESSORIES...
  • Page 14 GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 Blank page. 1–6...
  • Page 15: Room Requirements

    GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 CHAPTER 2 – ROOM REQUIREMENTS SECTION 1 ENVIRONMENTAL REQUIREMENTS/LIMITATIONS 1–1 Room climate TABLE 2-1 ENVIRONMENTAL REQUIREMENTS/LIMITATIONS HUMIDITY TEMPERATURE ALTITUDE Storage Storage In-Use In-Use In-Use Storage (equipment packed) (equipment packed) Min. Max. Min.
  • Page 16 GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 ILLUSTRATION 2-1 BASIC SYSTEM Configuration 1 3000 10' 1200 3.94' 600 (min.) 1.97' 600 (min.) 2' (min.) Configuration 2 3500 11.5' 1500 2.14' 0 1' * 600 mm (2'-0") is the absolute minimum access width...
  • Page 17: Structural Requirements

    GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 SECTION 2 STRUCTURAL REQUIREMENTS THE GANTRY MUST BE ANCHORED TO THE FLOOR IF STEREOTIX 2 IS USED. WARNING 2–1 Floor mounting The stand column is placed directly om‘n the floor. The floor must be stable and flat, and...
  • Page 18: Electrical Requirements

    GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 SECTION 3 ELECTRICAL REQUIREMENTS 3–1 Line voltage specifications Single–phase input voltage: - 200/208/220/240/380/415 V (±10%). 3–2 Line frequency specifications 50 or 60 Hz (±1 Hz). 3–3 kVA load characteristics 8.5 kVA for exposures up to 4.5 s.
  • Page 19: Line Supply Cable

    GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 3–5 Line supply cable The line supply cable comprises of two supply wires and a ground cable (i.e. 3 x AWG 10 (5.32 )) with the following actual/usable lengths: - Total length = 7 m (23'), - Usable length = 6.5 m (21'-4").
  • Page 20: Room Distribution

    GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 NEC 1993 Section 517–73 (a) Item 1: The ampacity of supply branch circuit conductors and the current rating of overcurrent protective devices shall not be less than 50 percent of the momentory rating or 100 percent of the long-time rating, whichever is greater.
  • Page 21: Iec60601-1-2 Electromagnetic Standards Compliance

    GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 SECTION 4 IEC60601-1-2 ELECTROMAGNETIC STANDARDS COMPLIANCE 4–1 General This equipment complies with the IEC60601-1-2 Edition 2 EMC standard for medical devices. The Senographe Equipment or System is suitable for use in electromagnetic environments as defined in the limits and recommendations given in the following tables: Emission Compliance level and limits (Table 2–1).
  • Page 22: Electromagnetic Immunity

    GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 4–3 Electromagnetic Immunity The Senographe is suitable for use in the specified electromagnetic environment. The purchaser or user of the Senographe should ensure that it is used in an electromagnetic environment as described below: TABLE 2–2...
  • Page 23 GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 TABLE 2–3 ELECTROMAGNETIC IMMUNITY –PART 2 Immunity Compliance Electromagnetic Environment Test 60601-1-2 Level Test Level ] 3 V At this disturbance level the monitor image may present some slight flicker. If this occurs, the monitor may be removed from...
  • Page 24 GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 4–4 Recommended Separation Distances for Portable and Mobile RF Communications Equipment IEC 60601-1-2 TABLE 2–4 RECOMMENDED SEPARATION DISTANCES Frequency of 150KHz to 26 26 MHz to 80 MHz 80 MHz to 800 800 MHz to 2.5...
  • Page 25 GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 The Senographe is primarily intended for use in non–domestic environments, and not connected directly to the public mains supply network. It is primarily intended for use in environments (such as hospitals) with a dedicated supply system, and in an X–ray shielded room.
  • Page 26 GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 SECTION 5 INSITE CONNECTION The Senographe DMR will use the Generic EDM (Equipment Diagnostic Monitor) when it is available (1st quarter 1996). The Generic EDM is supplied as a Service option, in a separate cabinet.
  • Page 27 GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 CHAPTER 3 – PRODUCT PHYSICAL CHARACTERISTICS 3–1...
  • Page 28 GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 ILLUSTRATION 3–1 GANTRY DIMENSIONAL 180° 180° 400 1.31' 150 (min.) 1200 3.94' 0.50' (min.) 0.74' VIEW A 1500 (min.) 4.93' (min.) 1500 (min.) 4.93' (min.) 3–2...
  • Page 29: Product Physical Characteristics

    GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 SECTION 1 DIMENSIONS TABLE 3–1 PRODUCT PHYSICAL CHARACTERISTICS DIMENSIONS WEIGHT ILLUSTRATION PRODUCT/ Length Width Height kg (pounds) COMPONENT mm (feet) mm (feet) mm (feet) Gantry 1200 (3.94’) (1.77’) 1710/2410 (5.61/7.91’) (617.3) 3–1 Console (1.81’)
  • Page 30 GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 ILLUSTRATION 3–2 GENERATOR DIMENSIONAL 1.84' 431 1.41' 295 0.97' 1.81' 1. Free volume for extra-cable length storage 0.5 m 3–4...
  • Page 31 GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 ILLUSTRATION 3-3 DIMENSIONS OF RAD SHIELD SCREEN WITH CONTROL CONSOLE 28" 3.20" 275 11" 21.5" 3–5...
  • Page 32 GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 ILLUSTRATION 3-4 DIMENSIONS OF OPTIONAL SCREEN 28" 3–6...
  • Page 33 GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 ILLUSTRATION 3-5 DIMENSIONS OF RAD SHIELD SREEN AND OPTIONAL SCREEN 1470 58.8' 275 11" 3–7...
  • Page 34 GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 ILLUSTRATION 3–6 DIMENSIONS OF RAD SHIELD SCREEN AND TWO OPTIONAL SCREENS 2240 89.6' 275 11" 3–8...
  • Page 35 GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 ILLUSTRATION 3-7 DIMENSIONS OF ACCESSORIES STORAGE DEVICE 900 2.95' 0.5 m 3–9...
  • Page 36 GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 ILLUSTRATION 3-8 DATAFLASH DIMENSIONS 3–10...
  • Page 37: Mounting Requirements

    GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 SECTION 2 MOUNTING REQUIREMENTS Given that the Senographe DMR has a modular design, its optimal use depends on the relative position of the modules (see the Installation Plan prepared for each site).
  • Page 38 GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 ILLUSTRATION 3-9 ISODOSE CURVE 50MSEV Section at 1,5m from the floor 40Kv, 4000mAs in 1 hour, 2000h/year, Mo track, large focus, Al filter, exposure on plexiglass insert (3 cm) 3–12...
  • Page 39: Additional Planning Aids

    GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 CHAPTER 4 – ADDITIONAL PLANNING AIDS 4–1...
  • Page 40 GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 ILLUSTRATION 4–1 INTERCONNECTION SCHEMATIC A = Input voltage = 480 V single–phase. B = Input voltage = 200/208/220/240/380/415 V single–phase. CUSTOMER GEMS–E SUPPLIED SUPPLIED DATAFLASH (OPTION) STEREOTIX 2 (OPTION) CONTROLLER Dia = 35 mm 1.38”...
  • Page 41: Product Shipping Information

    GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 SECTION 1 PRODUCT SHIPPING INFORMATION AIR AND ROAD SHIPMENT PRODUCT/ WEIGHT HEIGHT X WIDTH X DEPTH METHOD OF SHIPMENT COMPONENT POUNDS (daN) 6'36 X 3'. X 7'. SENO DMR 1476 (670) PALLET (1.94 X 0.90 X 2.14)
  • Page 42 GE Healthcare SENOGRAPHE DMR+ REV 1 pim 5128706–1–100 Blank page 4–4...