System Power Management; Rechargeable Battery Pack - GE VIVID S60N Basic Service Manual

Hide thumbs Also See for VIVID S60N:
Table of Contents

Advertisement

P R E L I M I N A R Y
D
FR091521, R
IRECTION
5-11-2

System Power Management

The System Power Management is responsible for operating the system in different power states, in
accordance with the power condition and User request.
5-11-3

Rechargeable Battery Pack

The rechargeable Battery pack is a smart battery device acting as a UPS and communicating with the
Vivid™ S60/Vivid™ S70 Ultrasound Scanner system over the SMBus.
The Battery pack contains protection circuitry, and a fuel gauge IC with internal memory to store pack
configuration information, measurement calibration, coefficients, manufacturer's information, and
chemistry data.
NOTE:
The Battery prevents uncontrolled system shut-down in the event of power loss, or if there is a need to
transport the system without having complete Power Off/On cycle.
In addition, the Battery enables fast return to scanning when AC power is resumed.
NOTE:
The system cannot scan while AC power is not present.
5-26
1
EVISION
Section 5-11 - System Power Distribution
V
™ S60N/V
™ S70N P
IVID
IVID
S
M
ROPRIETARY
ERVICE
ANUAL