Chapter 4 Electrical requirements

The purpose of this chapter is to ensure that the product is properly powered and grounded, thus
ensuring the proper operation of the product installed. The information in this chapter should be
adhered to, unless there are written deviations approved by GE Healthcare.
This chapter gives the requirements for powering and grounding the system. If these power and
grounding instructions are not adhered to, proper operation cannot be guaranteed. Any cost
associated and found to be a result of non-conformity may be the responsibility of the institution
and/or their contractor.

Section 4.1 Power requirements

4.1.1 Electrical requirements

Figure 4-1 Electrical outlet
3
EVISION
Customer-supplied electrical outlet (United States style; see
In regions that use an electrical plug other than the style used in the United States, the
installation process includes removing the electrical plug and replacing it with the appropriate
hospital-grade plug as dictated by local regulations. Detailed instructions are included in the
Optima XR200amx/Optima XR220amx Installation Manual.
System input voltage: 100 VAC or 120 VAC or 220/240 VAC
System input current (maximum input current during charging):
-
12 A at 110 V
-
7 A at 220 V
System input frequency: 50/60 Hz
Maximum 1 cycle in-rush current: 75 A peak
The system requires an AC outlet with a ground terminal.