Force Calibration And Bump; Zero Calibration; Capture Real Time Reading - Honeywell BW Flex Series User Manual

Portable multiple gas detector
Table of Contents

Advertisement

Force Calibration and Bump

Force Calibration has a higher priority than Force Bump. If you perform Force Calibration, there
is no need to execute the Force Bump again.
You can execute a Force Calibration/Bump via four methods: IntelliDoX docking module, SSDC,
DC app, and detector's menu.
Force Calibration/Bump via the IntelliDoX is executed automatically by the docking module.
To execute the Force Calibration/Bump via the SSDC, connect the detector to a computer and
follow SSDC onscreen instructions.
To execute the Force Calibration/Bump via de DC app, pair the detector to the mobile phone
and follow onscreen instructions.
To execute the Force Calibration/Bump via the detector's menu, enter the calibration/bump
menu and when Cal Now/Bump Now is displayed, press and hold the button and within 60s you
can enter to the flow of cal/bump.

Zero Calibration

Over time and through use, the sensor baseline at zero exposure may drift from the
manufacturer's baseline. For optimal performance Honeywell recommends that you zero the
sensors periodically. Only zero calibrate the sensor in fresh air.
1. Double press the button to enter the menu.
2. Single press to switch to ZERO CAL.
3. Press and hold the button to execute the Zero Calibration.
Zero calibration starts automatically, and the sensor LEDs light clockwise in blue.
After Zero calibration passes, "ZERO PASS" is displayed, the sensor LEDs light green for 5
seconds, the sound alarm beeps, and then the detector is back to the regular mode.

Capture Real Time Reading

1. Pair your BW Flex Series with a mobile device.
2. In your mobile device, open the Device Configurator app.
3. Tap Menu
4. Tap Measurements
5. Tap Start Recording.