Honeywell BW Flex Series User Manual page 16

Portable multiple gas detector
Table of Contents

Advertisement

Calibrate the detector if the gas readings varies during start up.
l
Calibrate the sensor before defining the alarm setpoints.
l
Calibrate only in a safe area that is free of hazardous gas in an atmosphere of 20.9% oxygen.
l
Do not calibrate the detector during or immediately after charging is complete.
l
The oxygen sensor can be automatically calibrated each time upon activation (if this feature
l
is enabled). Activate the detector in a normal (20.9%/20.8% oxygen) atmosphere.
Allow the detector to stabilize for 1 minute after activation before performing a calibration or
l
bump test.
If a certified calibration is required, contact Honeywell.
l
BW Flex Series
15
User Manual

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Bw flex-iBw flex4

Table of Contents