Troubleshooting

Problem
"Battery 0%" message is
displayed
Error 1006
Error 1007
Error 1008
Error 3001
Error 4004
Error 4006
Need Force Bump. "Bump
Now" message is displayed.
Need Force Calibration
"Cal Now" message is
displayed.
Detector alarms after start-
up sequence
Detector does not respond
when button is pressed
Detector does not respond
when button is pressed
BW Flex Series
Cause
Depleted battery
Temperature sensor fail
Data flash fail
BLE fail
RTC fail
The sensor is in the wrong
slot.
Sensors fail or no
communication
Bump overdue and must
carry out bump testing
before use.
Calibration overdue and
must carry out calibration
testing before use.
Sensor not stabilized
Sensors require calibration
The battery state is critically
low, or the battery is
depleted.
Detector is performing
operations that do not
require user input.
44
Solution
Charge the rechargeable
battery pack
Replace PCBA
Replace PCBA
Replace PCBA
Replace PCBA
Correct the sensor position.
Replace the sensor or the
PCBA
Hold the button for 3 seconds
or connect to DC/SSDC or
insert to IntelliDoX to start the
bump testing; otherwise, the
detector will auto power off
after 60 secs.
Hold the button for 3 seconds
or connect to DC/SSDC or
insert to IntelliDoX to start the
Calibration; otherwise, the
detector will auto power off
after 60 secs.
O
sensor: Wait for at least 10
2
min before power on.
For the NDIR-CH
sensor,
4
please allow 5 minutes
following startup for the
sensor to warm-up before
attempting to calibrate
Charge the rechargeable
battery pack
Button operation restores
automatically when the
operation ends.