MAN0923_Issue 3_04/15
Error
Message
Code
83
Kill Session denied Configuration
Backup / Restore in Progress
84
Kill Session denied CCB Firmware
update in Progress.
85
The controller needs a restart to
proceed further. Please power cycle
the controller and try again.
86
System is running in test mode.
87
Not used
88
Operation failed; Save Calibration
Failed.
89
Operation failed; Channel Wiring
Fault.
90
Operation failed; Bead Fault.
91
Operation failed; system is running
in test mode.
92
Calculation error
9. Problem Solving
Possible Cause
User has attempted to
end a session for the PC
configuration SW while
a CCB configuration
restore is in progress
from the local
touchscreen
User has attempted to
end a session for the PC
configuration SW while a
CCB firmware download
is in progress from the
local touchscreen
A system error has
occurred that can only
be cleared by a power
cycle
One of the three test
modes is active, and
other operations
are restricted or not
permitted
Touchpoint Pro was
unable to save the
sensor calibration
Touchpoint Pro cannot
detect that a sensor
is connected to the
channel
Touchpoint Pro cannot
read the bead current
from the connected mV
sensor
One of the three test
modes is active, and
other operations
are restricted or not
permitted
A calculation error
has occurred during
calibration, for example
Zero and SPAN
calibration done with the
same gas input
140
Remedial Action
Retry after the configuration restore operation is
completed
Retry after the firmware download operation is
completed
Please cycle the power to the Touchpoint Pro system
1. From the System Status screen select the
Extras icon then Diagnostics
2. Press the Menu icon, and from the popup menu
choose to stop the test mode that is active.
Please retry the operation.
Check all field wiring to the sensor and correct any
problems
Check all field wiring to the sensor. It is possible
that the sensor has failed, please refer to the
manufacturer's instructions
1. From the System Status screen select the
Extras icon then Diagnostics
2. Press the Menu icon, and from the popup menu
choose to stop the test mode that is active
Retry the calibration. Ensure that the values for zero
and span gas are entered correctly.
Touchpoint Pro
Operating Instructions