Adjustments / Programs

Dose-area product measurement device "PTW"

NOTE

Parameter "CONSTANT (Qt * Kg)" = T

Read and record the T
measurement device.

Programming

Connect the service PC to the serial interface of the SIREMOBIL Iso-C.
Start the service program.
Select the Adjustment, Parameter menu.
In the Combo box "Parameter Groups:", select "Diamentor" .
Click "Get from Unit".
The Parameter "1. Diamentor Existing" must be set to YES .
Select parameter "2. Constant (Qt * Kg)".
In "Value Actual:" field, enter the parameter T
Click "Set Value".
Click "Put to Unit".
Click "Save to File".
NOTE
© Siemens, 2008
For internal use only
This adjustment is used to calibrate the dose-area product
measurement device (PTW) of the SIREMOBIL Iso-C display.
w
value shown on the electronics unit of the dose-area product
w
You must exit the service program with "Logoff" and "Quit"
to store the modified parameters to disk as a new file.
SPR2-230.840.01.01.02