Video Switcher - Siemens SIREMOBIL Iso-C Troubleshooting Manual

Table of Contents

Advertisement

Replacing boards and components

Video switcher

Fig. 35:
NOTE
NOTE
Switch the system off.
Loosen the four fastening screws of the securing bracket
Loosen the two clamping screws, but do not unscrew them completely
((2/Fig. 35 / p.
Tip the holding plate with the video switchers into the monitor trolley compartment (see
(Fig. 36 / p.
Unplug the connectors of the video switcher to be replaced.
Pull the video switcher to be replaced out of the holding plate from the side.
Press the green button on the video switcher.
Push the new video switcher into the holding plate and reconnect all connectors.
Screw the holding plate back onto the monitor trolley.
Tighten the two clamping screws again.
Switch the system on.
After the system boots, the SIMOMED HM Monitor (reference monitor) should display
the MEMOSKOP reference image. If the boot process of the 3D reconstruction PC or
the syngo interface appears, the green button of video switcher 1 is in the wrong posi-
tion and needs to be switched over.
Perform the IQ quick test.
© Siemens, 2008
For internal use only
Only applies to SIREMOBIL Iso-C 3D.
Video switcher 1
video signal between the MEMOSKOP and the 3D recon-
struction PC. Video switcher 2
route the video signal to the video printer.
121)).
121)).
((3/Fig. 35 / p.
SPR2-230.840.01.01.02
08.08
Fig. 36:
((3/Fig. 35 / p.
121)) is used to switch the
((1/Fig. 36 / p.
((1/Fig. 35 / p.
121)) points to video switcher 1.
Page 121 of 148
CS PS SP
121
6.21
121)) is used to
121)).
SIREMOBIL Iso-C

Advertisement

Table of Contents
loading

Table of Contents