Adjustments / Programs
Potentiometer
Fig. 6
Calibrating the area dose product measurement device
NOTE
CONSTANT parameter (Qt * Kg) = T
•
Read and record the
measurement device.
Programming
•
Connect the service PC to the serial interface of the SIREMOBIL Iso-C.
•
Start the service program.
•
Select the Adjustment, Parameter menu.
•
In the Combo box "Parameter Groups:", select "Diamentor" .
•
Click on "Get from Unit".
•
The Parameter "1. Diamentor Existing" must be on YES .
•
Select parameter "2. Constant (Qt * Kg)" .
•
In "Value Actual:"field, enter the parameter
•
Click on "Set Value".
•
Click on "Put to Unit" .
•
Click on "Save to File".
NOTE
Siemens AG
Medical Solutions
This adjustment is used to calibrate the area dose product mea-
surement device of the SIREMOBIL Iso-C display.
T
value shown on the electronics unit of the area dose product
w
You must exit the service program with "Logoff" and "Quit" to
store the modified parameters to disk as a new file.
SPR2-230.061.01
Rev. 13
w
T
.
w
Page 21 of 30
12.04
CS PS 24
5 - 21
5
5
5
SIREMOBIL Iso-C
System Manual