Check the camera rotation and, if necessary, adjust it.

Concluding work

Reinstall the I.I. on the C-arm flange. See
Check the collimator display on the monitor and, if necessary, adjust it.
Perform the I.Q. Quick test.
Check the setting of the laser light localizer, if present. If necessary, adjust it.
If the "MODULARIS" system is present, perform the isocenter check.
Execute the country-specific acceptances (§16 Partial acceptance test, DHHS, etc.).
SIREMOBIL Iso-C
Replacing boards and components
SPR2-230.840.01.01.02
08.08
(I.I. service position / p. 25)
Page 98 of 148
CS PS SP
0
© Siemens, 2008
For internal use only