74
Overview of MEMOSKOP C / C-E / C-SUB / C & MOD / C-SUB &
MOD Parameters
User Setup
User Setup is selected with the CTRL + U key combination.
The available functions are described in the SIREMOBIL Iso-C Operating Instructions.
Technical Setup
Technical Setup is selected with the CTRL + T key combination.
Enter the password and confirm with [RETURN]
The following functions can be selected depending on the type of MEMOSKOP:
Function:
Test pattern generation
Video frequency
Number of monitors
Hardcopy setup
Edge 1 coef set
Edge 2 coef set
Motion detection setup
Circular mask setup
Move text positions
Date / Time format
System configuration
Keyboard
Graphics level
Display of real time clock
SIREMOBIL Iso-C
Comments:
Not for service. Select the necessary test images on the ser-
vice PC.
The currently programmed frame rate is shown in the paren-
theses to the right. After downloading the MEMOSKOP or
after replacement, the MEMOSKOP will start up at the
default video frequency (50 Hz). If the monitors are config-
ured for a different video frequency, you must perform "blind"
configuration for the correct video frequency. Refer to "Con-
figuring video frequency".
Allows you to program the number of monitors used (1 or 2)
You can configure the connection for hardcopy cameras.
Check whether 5000 is programmed for "4 Max Hardcopy
time" and if not, program this value.
Allows configuration of edge enhancement. Check whether
3 is programmed and if not, program this value.
Allows Configuration of edge enhancement. Check whether
8 is programmed and if not, program this value.
Configures the motion detector. Check if at MD 1: Threshold
100 and at MD 2:Threshold 120 is programmed, if not, pro-
gram these values.
Allows you to configure the diameter of the circular mask.
Allows you to configure the position of superimposed text.
European: DD MM YY / HH MM
US:
Allows you to read out the current MEMOSKOP configura-
tion.
Allows you to program the corresponding keyboard layout
Universal / Roman.
Level (brightness) of superimposed text. (Default: 128)
Yes: Real time clock: date and time are displayed according
to the user setup programming for MEMOSKOP. The real
time clock is active. NO: The real time clock is switched off.
The user setup programming will not affect it.
SPR2-230.840.01.01.02
08.08
Adjustments / Programs
MM DD YY / HH MM
Page 74 of 148