Checking the operating voltages

I.I. voltages

Voltages E1 / E2 / E3 and A may be taken from the I.I. test protocol and checked or adjusted
against the control test points listed in the table.
Voltage
E1
E2
E3
30 kV, anode
© Siemens, 2008
For internal use only
Ground
Test point
point (0V)
UE1
UE2
UI 15
UI 30
SPR2-230.840.01.01.02
08.08
Potentiome-
Potentiome-
ter for full
ter for zoom
format
format
P10
P11
P6
P7
P2
P3
P1
P1
Page 31 of 148
CS PS SP
31
3.1
Voltage
divider
ratio
1:1
1:1
1:10000