Connecting laser camera

Power cable

Connect the power cable of the video separator device at terminals X2.1 and X2.2
(Fig.1).
Connect the protective conductor to the protective conductor terminal strip (yellow/
green) (Fig.1).
Route the power cable to the video separator device and connect it.

Video cables

When a multi-format camera or video printer is present
Disconnect the 75 Ohm terminating resistor from the video output of the multi-format
camera or disconnect the switchable 75 Ohm termination of the video printer.
Connect the video cable to the video output of the multi-format camera or the video
printer, route it to the video input of the optical video separator device, and connect it.
Route the second video cable from the output of the optical video separator device to the
insulated, built-in video socket and connect it.
Without a multi-format camera or video printer
Connect the video cable at the MEMOSKOP "Out3" connector.
Route the video cable to the video input of the optical video separator device.
Route the second video cable from the output of the optical video separator device to the
insulated, built-in video socket and connect it.
When a video recorder is present
Disconnect the 75-Ohm terminating resistor of the video recorder.
Connect the video cable to the video recorder output and route it to the video input of the
optical video separator device.
Route the second video cable from the output of the optical video separator device to the
insulated, built-in video socket and connect it.