Connecting laser camera
Power cable
•
Connect the power cable of the video separator device at terminals X2.1 and X2.2
(Fig.1).
•
Connect the protective conductor to the protective conductor terminal strip (yellow/
green) (Fig.1).
•
Route the power cable to the video separator device and connect it.
Video cables
When a multi-format camera or video printer is present
•
Disconnect the 75 Ohm terminating resistor from the video output of the multi-format
camera or disconnect the switchable 75 Ohm termination of the video printer.
•
Connect the video cable to the video output of the multi-format camera or the video
printer, route it to the video input of the optical video separator device, and connect it.
•
Route the second video cable from the output of the optical video separator device to the
insulated, built-in video socket and connect it.
Without a multi-format camera or video printer
•
Connect the video cable at the MEMOSKOP "Out3" connector.
•
Route the video cable to the video input of the optical video separator device.
•
Route the second video cable from the output of the optical video separator device to the
insulated, built-in video socket and connect it.
When a video recorder is present
•
Disconnect the 75-Ohm terminating resistor of the video recorder.
•
Connect the video cable to the video recorder output and route it to the video input of the
optical video separator device.
•
Route the second video cable from the output of the optical video separator device to the
insulated, built-in video socket and connect it.
Siemens AG
Medical Engineering
Register 3
RXR2-130.033.02
Rev. 02
Page 5 of 8
03.97
TD SP 1
5
1
1
1
1
1
SIREMOBIL Compact