Physiologic Alarms
Alarm Text
** MULTI ST Lx, Ly
** NURSE CALL
** SpO
T yyy > xxx
2
** SpO
T yyy < xxxx
2
** ST lead > xxx
** ST lead < xxx
* HR yyy > xxx
* HR yyy < xxx
5-6
Alarms
Physiologic (Patient) Alarms
Priority
Condition
Yellow
Two ST leads (Lx and Ly) exceed alarm
limit elevation or depression for > 60
seconds (EASI mode or when selected).
Yellow
The Telemetry button on the transceiver
has been pressed (and the system is
configured to alarm and the Telemetry
button is on.
Also initiated if the installation includes a
paging system and the Information Center
is configured for paging upon receipt of
Nurse Call signal.
For IntelliVue Information Centers
Note—
Release J and later, Nurse Call is a
yellow, one star (*) alarm.
Yellow
High SpO
SpO
value (yyy) greater than high SpO
2
limit (xxx).
Yellow
Low SpO
SpO
value (yyy) less than low SpO
2
(xxx).
Yellow
STlead is greater than the Elevation limit.
Yellow
STlead is less than the Depression limit.
Yellow
Heart Rate (yyy) greater than the upper
Heart rate limit (xxx).
Yellow
Heart Rate (yyy) lower than the lower
Heart Rate limit (xxx).
.
2
.
2
Source
ST/AR ST
Clinician-
initiated at
transceiver
or by
paging
subsystem
SpO
2
2
SpO
2
limit
2
ST/AR
ST/AR
ST/AR
Basic &
Enhanced
Arrhythmia
ST/AR
Basic &
Enhanced
Arrhythmia