Philips IntelliVue TRx Instructions For Use Manual page 190

For the its4840a/its4850a intellivue telemetry system
Hide thumbs Also See for IntelliVue TRx:
Table of Contents

Advertisement

Cleaning and Sterilization
Test Procedure
11-10
Maintenance, Cleaning & Troubleshooting
Caution
You must perform this test each time you put a transceiver through the EO
sterilization process.
This test allows you to verify that patient information for both ECG and SpO
you are monitoring pulse oximetry) appear at the Information Center and at the
bedside. You can use this procedure with a Patient Simulator.
This test assumes that the telemetry system and Information Center are
Note—
fully installed, and that you have performed the procedure to learn the
transceiver identity code.
Test the transceiver by performing the following steps. If the test indications do
not appear, refer to your service provider.
Step
Action
1
Perform a mechanical inspection of the transceiver (connectors,
battery door opening and closing, Telemetry and Check buttons).
2
At the Information Center, select the telemetry bedside you are
testing.
(if
2