Alarm Text
***ASYSTOLE
*** BRADY yyy < xxx
*** DESAT
*** TACHY yyy > xxx
*** V-FIB/TACH
*** V-TACH
NBP alarms are listed in the TeleMon Instructions for Use and the Information
Center Online Help.
In the following table, Red (***) alarms are listed alphabetically, followed by
the Yellow (**) alarms, and the Yellow (*) arrhythmia alarms.
Physiologic (Patient) Alarms
Priority
Red
Red
Red
Red
Red
Red
Condition
Asystole.
No QRS for 4 consecutive seconds
Extreme Bradycardia.
Heart Rate (yyy) less than Extreme Brady
limit (xxx)
Very Low SpO
Saturation.
2
SpO
value below Desaturation limit
2
Desat limit is set 10 points below
Note—
low limit.
Extreme Tachycardia.
Heart Rate (yyy) greater than Extreme
Tachy limit
Ventricular Fibrillation.
Fibrillatory wave (sinusoidal wave
between 2-10 Hz) for 4 consecutive
seconds
Ventricular Tachycardia.
Consecutive PVCs greater than or equal to
V-Tach Run limit and Heart Rate greater
than V-Tach limit (xxx)
Physiologic Alarms
Source
ST/AR
Basic &
Enhanced
Arrhythmia
ST/AR
Enhanced
Arrhythmia
SpO
2
ST/AR
Basic &
Enhanced
Arrhythmia
ST/AR
Basic &
Enhanced
Arrhythmia
ST/AR
Basic &
Enhanced
Arrhythmia