Measuring Spo; Spot Check Measurement - Philips IntelliVue TRx Instructions For Use Manual

For the its4840a/its4850a intellivue telemetry system
Hide thumbs Also See for IntelliVue TRx:
Table of Contents

Advertisement

Measuring SpO

Spot Check
Measurement
2
Warning
Warning
Removal of the SpO
a "No Sensor" technical alarm. There is no technical alarm for a "No
Sensor" condition in Spot Check mode.
Warning
Warning
If you measure SpO
pulsatile SpO
technical alarm can occur. If the monitor is configured to
2
suppress this alarm, there can be a delay of up to 60 seconds in indicating
critical patient status, such as sudden pulse loss or hypoxia.
SpO
measurements can be made manually on an as-needed basis in Spot
2
Check mode, or continuously in Continuous mode, depending on the transceiver
configuration. While operating in Continuous mode, you can also measure
pulse, and display the pleth wave at the Information Center. The SpO
parameter is turned on by inserting/removing the sensor cable into the
transceiver, or by control from the Information Center. SpO
consumes considerable electrical energy. The battery power must be at least
25% full in order to make SpO
Setting the mode at Spot Check or Continuous is done at the Information Center
(see "Patient Configurable Settings in Telemetry Setup" on page 9-4) or at
TeleMon.
When the transceiver is configured for Spot Check measurement, use the
following instructions to take an individual, manual SpO
transceiver.
You can also initiate a Spot Check measurement in the Patient Window
Note—
at the Information Center by selecting the STAT SpO
Configurable Settings in Telemetry Setup" on page 9-4).
sensor during Continuous SpO
2
on a limb that has an inflated NBP cuff, a non-
2
measurements.
2
Measuring SpO
monitoring results in
2
2
monitoring
2
reading from the
2
icon (see "Patient
2
SpO
Monitoring
2
2
8-15