Connecting the SpO
2
Connecting the SpO
8-14
SpO
Monitoring
2
Cable
Cable
2
Step
Action
1
Connect the sensor cable to IntelliVue TRx
•
Connect reusable sensors directly into the transceiver.
•
Connect disposable sensors into the adapter cable, then
connect the adapter cable to the transceiver. Remove the
protective backing.
2
Adjust SpO
Controls in the Patient Window" on page 9-2).
3
Make other adjustments in the Telemetry Setup Window (see
"Patient Configurable Settings in Telemetry Setup" on page 9-4).
Caution
Extension cables: The use of extension cables presents a tripping hazard. Use
appropriate care when they are in use.
Electrical Interference: Position the sensor cable and connector away from
power cables, to avoid electrical interference.
alarms in the Patient Window (see "Telemetry
2
+
.