Spo 2 - Philips IntelliVue TRx Instructions For Use Manual

For the its4840a/its4850a intellivue telemetry system
Hide thumbs Also See for IntelliVue TRx:
Table of Contents

Advertisement

SpO
2
Parameter
SpO
Measurement
2
Range (Calibration and
Display)
SpO
Accuracy
2
SpO
Resolution
2
SpO
Numerics -
2
Averaging
SpO
& Pulse Numerics
2
- Update Rate
Pleth Wave- Sampling
Rate
Technical Alarms
(INOPs)
Specification
0 to 100%
See table following.
1%
10 seconds
The update rate for the SpO
Note—
oximetry value and pulse rate is typically 1
second. This can be extended to a max. 60 s
when NIBP is measured on the same limb, with a
corresponding INOP message after a max. of 30
s, indicating that the displayed values are not
current values.
The effect of SpO
pulse oximetry on data
2
averaging is internally controllable by the
transmitter, with no user controls.
Transmitted once per second.
125 sps
Triggered if the sensor is disconnected, if a pulse
is not detected, if the signal is noisy, if light
interference is detected, if the sensor is defective,
if the measurement is erratic, or if the module is
malfunctioning