SpO
2
Parameter
SpO
Measurement
2
Range (Calibration and
Display)
SpO
Accuracy
2
SpO
Resolution
2
SpO
Numerics -
2
Averaging
SpO
& Pulse Numerics
2
- Update Rate
Pleth Wave- Sampling
Rate
Technical Alarms
(INOPs)
Specification
0 to 100%
See table following.
1%
10 seconds
The update rate for the SpO
Note—
oximetry value and pulse rate is typically 1
second. This can be extended to a max. 60 s
when NIBP is measured on the same limb, with a
corresponding INOP message after a max. of 30
s, indicating that the displayed values are not
current values.
The effect of SpO
pulse oximetry on data
2
averaging is internally controllable by the
transmitter, with no user controls.
Transmitted once per second.
125 sps
Triggered if the sensor is disconnected, if a pulse
is not detected, if the signal is noisy, if light
interference is detected, if the sensor is defective,
if the measurement is erratic, or if the module is
malfunctioning
Safety Standards & Specifications
Measurement Specifications
pulse
2
12-17