Transceiver Operation when Connected to TeleMon
Transceiver Operation when Connected to TeleMon
9-16
Telemetry Functions at the Information Center & TeleMon
When connected to TeleMon, the following differences in transceiver operation
will be in effect.
The transceiver must have good batteries installed before being
Important—
connected to TeleMon.
Alarms
•
Alarm Suspend/Pause and Standby can be initiated only from TeleMon,
not from the transceiver or the Information Center. The alarm suspend
indicator on the transceiver will accurately reflect the current state of
alarm pause.
ECG Operation
•
Vb, the second V-lead in 6-wire lead-set, is not supported.
•
After a change in leadset, the TeleMon returns to the default ECG
settings, and arrhythmia relearn occurs automatically. Be sure to check the
monitoring leads after you switch leadsets.
SpO
Operation
2
•
SpO
is always in continuous mode.
2
•
Changes in SpO
disconnected from TeleMon. Mode settings are defined in the following
table:
Mode Selected at TeleMon
Continuous
5-min.
1-min.
Manual
•
If the sensor is disconnected from the transceiver during SpO2 monitoring
at TeleMon, a No Sensor technical alarm will be displayed at the
Information Center. NO VISUAL OR AUDITORY INDICATORS FOR
THE NO SENSOR ALARM WILL BE ACTIVE AT TELEMON.
mode do not take effect until after the transceiver is
2
Mode at Disconnected Transceiver
Continuous
Continuous
Continuous
Spot Check